Appendix E: Serial Number Tag Description
Product Service 1-800-522-7658
57
Original instructions
D:
This fume hood includes
three
additional electrical receptacles,
individually wired to the field wired box and individually rated as 115V,
single phase, 60Hz, 20 Amps. Each duplex can be wired on a dedicated
circuit rated at 20A or the duplexes may be ganged together on the same
circuit with the total load of the ganged duplexes not exceeding the 20A
electrical service they are wired to.
E:
This fume hood includes
four
additional electrical receptacles,
individually wired to the field wired box and individually rated as 115V,
single phase, 60Hz, 20 Amps. Each duplex can be wired on a dedicated
circuit rated at 20A or the duplexes may be ganged together on the same
circuit with the total load of the ganged duplexes not exceeding the 20A
electrical service they are wired to.
F:
This fume hood includes
one
additional electrical receptacle, individually
wired to the field wired box and individually rated as 230V, single phase,
60Hz
, 10 Amps. Each duplex can be wired on a dedicated circuit rated at
10A or the duplexes may be ganged together on the same circuit with the
total load of the ganged duplexes not exceeding the 10A electrical service
they are wired to.
G:
This fume hood includes
two
additional electrical receptacles, individually
wired to the field wired box and individually rated as 230V, single phase,
60Hz
, 10 Amps. Each duplex can be wired on a dedicated circuit rated at
10A or the duplexes may be ganged together on the same circuit with the
total load of the ganged duplexes not exceeding the 10A electrical service
they are wired to.
H:
This fume hood includes
three
additional electrical receptacles,
individually wired to the field wired box and individually rated as 230V,
single phase,
60Hz
, 10 Amps. Each duplex can be wired on a dedicated
circuit rated at 10A or the duplexes may be ganged together on the same
circuit with the total load of the ganged duplexes not exceeding the 10A
electrical service they are wired to.
J:
This fume hood includes
four
additional electrical receptacles,
individually wired to the field wired box and individually rated as 230V,
single phase,
60Hz
, 10 Amps. Each duplex can be wired on a dedicated
circuit rated at 10A or the duplexes may be ganged together on the same
circuit with the total load of the ganged duplexes not exceeding the 10A
electrical service they are wired to.
K:
This fume hood includes
one
additional electrical receptacle, individually
wired to the field wired box and individually rated as 230V, single phase,
50Hz
, 10 Amps. Each duplex can be wired on a dedicated circuit rated at
10A or the duplexes may be ganged together on the same circuit with the
total load of the ganged duplexes not exceeding the 10A electrical service
they are wired to.
L:
This fume hood includes
two
additional electrical receptacles, individually
wired to the field wired box and individually rated as 230V, single phase,
50Hz
, 10 Amps. Each duplex can be wired on a dedicated circuit rated at
10A or the duplexes may be ganged together on the same circuit with the
total load of the ganged duplexes not exceeding the 10A electrical service
they are wired to.