Product Service 1-800-522-7658
18
Original instructions
C
HAPTER
4
P
ERFORMANCE
F
EATURES
A
ND
S
AFETY
P
RECAUTIONS
Performance Features:
The Protector Premier Laboratory Hood is designed to meet the needs of the
laboratory scientist at OSHA approved velocities as low as 60 feet per minute.
The laboratory fume hood has been designed to effectively contain toxic, noxious,
or other harmful materials when properly installed. A fully featured by-pass hood
with baffle and air foil, this hood maintains safe airflow. Optional energy saving
A-Style Combination Sash models are also available. The hood features by-pass
airflow design that promotes full containment as the sash is moved. Airflow is
diverted behind the front panel and under the air foil to help control fluctuations
in face velocity, which occur as the sash is closed.
1.
Unique sash provides maximum visibility of 37.5" high while
conserving energy by limiting sash travel to 28"
. Vertical-rising sash
may be raised from a closed to 28" operating height. The fume hood
features a containment-enhancing sash handle. Optional sash stops are
available to limit sash height and reduce energy usage.
2.
By-pass airflow design
ensures containment as the sash moves.
3.
Large usable interior work depth and interior height of 48"
provides
ample working space.
4.
Baffle
(not shown)
directs airflow to the rear of the interior to provide
efficient airflow. The baffle may be removed for cleaning purposes only.
5.
Exterior access cover plates
are removable for easy access to plumbing
valves and sash adjustment hardware when access through the sides is not
available.