PPENDIX
E
S
ERIAL
N
UMBER
T
AG
D
ESCRIPTION
Serial tag includes standard information with the following changes:
The portion of the tag for electrical information is labeled with the following
units: Volts AC, Phase, Cycle, AMP.
Each hood will include one of the following:
Volts AC
Phase
Cycle
AMP x
115
1
50/60
10X
115
1
50
10X
115
1
60
10X
230
1
50/60
5X
230
1
50
5X
230
1
60
5X
Where
may be a separate label indicating: Caution, See Manual.
The X above will be one of several possible alpha characters as defined in the
manual as follows:
X:
No additional duplexes on hood, noted amperage applies.
B:
This fume hood includes
one
additional electrical receptacle, individually
wired to the field wired box and individually rated as 115V, single phase,
60Hz, 20 Amps. Each duplex can be wired on a dedicated circuit rated at
20A or the duplexes may be ganged together on the same circuit with the
total load of the ganged duplexes not exceeding the 20A electrical service
they are wired to.
C:
This fume hood includes
two
additional electrical receptacles, individually
wired to the field wired box and individually rated as 115V, single phase,
60Hz, 20 Amps. Each duplex can be wired on a dedicated circuit rated at
20A or the duplexes may be ganged together on the same circuit with the
total load of the ganged duplexes not exceeding the 20A electrical service
they are wired to.