Chapter 3: Getting Started
Product Service 1-800-522-7658
13
Original instructions
Connecting to the Hood Exhaust System
WARNING: The weight of the exhaust ductwork system must
be supported independently of the hood superstructure. Do
not allow this weight to be supported by the hood structure as
damage to the hood may occur.
AVERTISSEMENT : Le poids du système d'aspiration de
ductwork doit être soutenu d'une manière indépendante de la
superstructure de la hotte. Au cas où ce poids est supporté par
la structure de la hotte, des dommages à la hotte peuvent
arriver.
The exhaust connection should be installed by a qualified
HVAC contractor
.
The exhaust connection on your hood has
been designed for 12" nominal pipe (12.75" OD) to allow for
minimum static pressure loss with proper transport velocities
away from the hood. Consult Labconco Customer Service
should you require help sizing your blower for the exhaust
volume and total system static pressure loss.
La connexion d'échappement devrait être installée par un
professionnel de CVC agrée. La connexion d'aspiration sur
votre hotte a été conçue pour un tuyau nominal de 12 pouces
(12,75 pouces de diamètre externe) afin d’avoir une perte
minimale de pression statique avec les correctes flux de
transport loin de la hotte. Consulter le Service Clientèle de
Labconco si la calibration de votre soufflerie pour le volume
d'aspiration et la perte de pression statique du système le
requiert.
The selected exhaust duct material should match the hood
procedures and chemicals used to ensure compatibility.
Le matériel de conduit d'aspiration choisi devrait être en
conformité avec les procédures de la hotte et les produits
chimiques qui sont utilisés pour garantir la compatibilité.
Connecting the Electrical Supply Source
to the Protector Fume Hood
Prior to connecting any electrical wiring to the fume hood structure, refer to the
hood identification plate for the proper electrical requirements of your specific
model.
!
!