Apture™ Portable  

Fuming Systems

 

 
 
 
 
 
 

Models 

3420000 
3420020 
3420022 
3420024 

 

 
 
 
 
 

Summary of Contents for CApture 3420000

Page 1: ...CApture Portable Fuming Systems Models 3420000 3420020 3420022 3420024...

Page 2: ...7 30 a m and 5 30 p m Central Standard Time Part 3421500 Rev C ECO L627 Warranty Labconco Corporation provides a warranty to the original buyer for the repair or replacement of parts and reasonable l...

Page 3: ...tem 5 Blower 6 Safety Precautions 6 8 CHAPTER 4 USING THE SYSTEM 9 Unpacking the System 9 Transporting the System 10 Component Identification 11 Hints for Handling CA 12 Preparing the System for Opera...

Page 4: ...APPENDIX B DIMENSIONS 24 APPENDIX C SPECIFICATIONS 25 APPENDIX D ACCESSORIES 26...

Page 5: ...of the development process The Portable Fuming System offers many unique features to enhance performance and flexibility To take full advantage of them please acquaint yourself with this manual and ke...

Page 6: ...part number 34252002 A supply of Rain X1 or Rain Block2 to treat the clear chamber before use Exhaust Requirements If you intend to connect the system to the optional Exhaust Connection Kit first exa...

Page 7: ...200 20 22 24 230 VAC 50 60 Hz 0 5 Amp Note An outlet with a circuit breaker rated at 10 amps for 115 volt models 5 amps for 230 volt models should be located as close as possible to the system Power c...

Page 8: ...nce of water specifically hydroxide ions forming long strong chains joining the bonded surfaces together CA is a tenacious adhesive particularly when used to bond non porous materials or those that co...

Page 9: ...It protects the heater filter module when the unit is being transported and allows an unobstructed view of the evidence during the fuming process While the clear chamber is resistant to many chemicals...

Page 10: ...ful rapid exothermic reaction The heat released may cause minor burns and if enough cyanoacrylate is used the reaction is capable of igniting the cotton product as well as releasing irritating vapor i...

Page 11: ...r is detected coming out of the granular activated carbon it needs to be replaced The system weighs approximately 22 lbs 10 kg Always follow safe lifting guidelines when transporting the unit When cle...

Page 12: ...conco s Product Service Department for questions on determining safe state of this product Verify protective earth ground continuity before using equipment No special ventilation requirements No hazar...

Page 13: ...n transit If the system is damaged notify the delivery carrier immediately and retain the entire shipment intact for inspection by the carrier Note United States Interstate Commerce Commission rules r...

Page 14: ...ne or more of the components listed for the system contact Labconco Corporation immediately for further instructions The system is shipped in the transport mode as shown in Figure 4 1 Transporting the...

Page 15: ...intenance Panel Thumbscrew Control Switch Heat Off Vent Circuit Breaker Main Power Switch Power Cord Connection Heater Receptacles 2 Leveling Foot Pin Leveling Foot Blower Exhaust Stainless Steel Face...

Page 16: ...their usage CA will react and polymerize on any surface or material that is wet or damp NEVER use the system on wet or damp evidence If you are trying to reuse CA heating pans discard them after the...

Page 17: ...ve the accessories and the foam storage insert from the bottom of the clear chamber 2 Put the leveling foot in the down position by pulling the pin out and sliding the foot down until the pin sets as...

Page 18: ...5 5 Place the evidence so that fumes can contact as much of its surface as possible If desired place a control fingerprint on a glass slide and set the slide into one of the slide slots in the face p...

Page 19: ...N Button SET Button 2 To check or change the temperature setting a Press SET SP will appear on the display b Press SET again The current temperature setting is displayed c The setting can be changed b...

Page 20: ...pment of fingerprints is 70 90 with the most high quality marks developing when relative humidity is approximately 80 3 It has been determined by calculation and proven by in house testing that one dr...

Page 21: ...to open 2 Open the heater door Place approximately 4 to 6 drops of CA glue in a heating pan and place it in one of the heater receptacles as shown in Figure 4 7 and immediately close the door Note Do...

Page 22: ...rom the stainless steel face plate 10 Wipe up any residue from the stainless steel face plate and the clear chamber 11 Unplug the unit Stow the foam storage insert and the accessories in the bottom of...

Page 23: ...he inside surface of the clear chamber is cleaned reapply either Rain X or Rain Block treatmentTM Monthly or more often as required Using a damp cloth or LabSolutions Glass Surface Wipes Labconco part...

Page 24: ...clear chamber Place the module with the stainless steel face plate down on a stable level surface as shown in Figure 5 1 2 Loosen the maintenance panel thumbscrew It is a captive screw and will be re...

Page 25: ...ge out until it clears the back edge of the heater filter module then pull it out of the flexible connector 5 Discard the used cartridge and install a replacement by reversing the removal steps 6 Ensu...

Page 26: ...2 drops of water in a heating pan and place on the heater Little or no fumes are being produced Heater temperature is too low Increase heater temperature Bottom of heater pan is dented bent Discard d...

Page 27: ...System Parts Item Quantity Part No Description 1 1 3421901 Clear chamber 2 1 3425202 Heating pans package of 250 not shown 3 1 3424500 Flexible Connector 4 1 3421801 Filter Cartridge 5 1 3422203 Shel...

Page 28: ...Product Service 1 800 522 7658 24 Appendix B Dimensions Clear Chamber Interior Dimensions 16 406mm H x 8 203mm W x 8 203mm D B 1...

Page 29: ...decreasing linearly to 50 relative humidity at 104 F 40 C Main supply voltage fluctuations not to exceed 10 of the nominal voltage Transient overvoltages according to Installation Categories II Overv...

Page 30: ...3425400 Allows the system to exhaust to a remote location or a nearby chemical fume hood Includes 10 1 2m long extended full length 4 100 mm diameter collapsible duct 1 lb 0 5kg Replacement Hook 3422...