Chapter 4: Using the System
Product Service 1-800-522-7658
18
5. Let the chamber vent for at least 3-5 minutes before opening the
chamber and retrieving the evidence.
6. If more evidence needs to be developed, turn the control switch to
“Heat” and go back to step #1 and repeat.
7. If no more evidence needs to be developed, push the control switch
to the center “OFF” position.
8. Remove any heating pans from the heater receptacles.
9. Remove any evidence hooks or shelves from the stainless steel
face plate.
10. Wipe up any residue from the stainless steel face plate and the
clear chamber.
11. Unplug the unit. Stow the foam storage insert and the accessories
in the bottom of the clear chamber.
12. Raise the leveling foot to the up position by pulling the pin out, and
sliding the foot up until the pin sets, as shown in Figure 4-4.
13. Place the heater/filter module into the clear chamber, ensuring that
is fits into the chamber completely.
14. Place the transport lid back on top of the stainless steel face plate,
and reengage the two clamps securing the transport cover to the
clear chamber. The unit is now ready for transport or storage.