Chapter 6:
Troubleshooting
Refer to the following table if the system fails to operate properly. If the
suggested corrective actions do not solve the problem, contact Labconco
for additional assistance.
PROBLEM
CAUSE
CORRECTIVE ACTION
Blower and controller
won’t turn on
Unit not plugged into
outlet
Plug the cabinet into appropriate
electrical service.
Circuit breaker
tripped
Reset circuit breaker.
Prints are not
developing properly
Insufficient humidity
in the clear chamber
Place 1-2 drops of water in a heating
pan, and place on the heater.
Little or no fumes are
being produced
Heater temperature is
too low.
Increase heater temperature
Bottom of heater pan
is dented/bent
Discard damaged pan, and use one
with flat bottom.
Heater temperature
not changing on
controller (“out”
indicator is on)
Heater /Over
temperature switch
failure
Contact Labconco Product Service
Department.
Heater temperature
not changing on
controller (out
indicator is off)
Controller is defective Contact Labconco Product Service
Department.
Fan does not run
Fan or 12 volt power
supply failure
Contact Labconco Product Service
Department.
Controller displays
“Er” or “oo” or “-”
Controller error
Contact Labconco Product Service
Department.