39
Drying Accessory Screen
Drying Accessory Sensor Table
– The middle section of the screen displays a list of
all the sensors that are available for the drying accessory. Sensors that are
disconnected will have three dashes (---) in place of a value.
Drying Accessory Status Box
– The status box is located directly below the sensor
table. The text on the left side will indicate the current operation mode of the tray dryer
(Off, Manual, or P
rogram details). When in manual mode, the word “Manual” and the
current shelf set point temperature value will be displayed in this location. If a program
is running the program name, step # and time remaining (TR) of the current step will be
displayed in this location. The right side of the status box will indicate if Data Logging is
currently ON or OFF.
Program
– Pressing this button will take you to the Programs screen where
you can create a new program or start, edit, view, copy and delete a saved
program.
Manual
– Use this button to enter a Shelf Set Point Temp and/or start the tray
dryer in manual mode.