25
Moving Between Menu Screens
You can move between menu screens by either touching the menu icons or swiping
across the screen from left to right or right to left.
The following system parameters are displayed in the center portion of the screen:
Indicator Bar Function
A thin blue indicator bar will appear under the icon of the menu screen that is currently
being displayed. The same type of indicator bar will appear above the operation buttons
to signify if each function is ON or OFF.
System Parameters Display
Line 1: Collector & Vacuum Heading
Describes the measurement that is displayed directly below the text.
Line 2: Collector Measurement
Displays the temperature of the collector coil in ⁰C or ⁰F, as selected in Settings. The
value displayed with be
RED
if the temperature is > -
40⁰C.
The value displayed will be
GREEN
if the temperature is ≤ -40⁰C.
Vacuum Measurement
Displays the freeze dryer vacuum level in mbar, Pa, or Torr, as selected in Settings. The
display will read “High” if the vacuum level is > 5mbar. If the vacuum level is ≤ 5mbar, a
numerical value will be displayed.
The value displayed will be
RED
if the vacuum is > 1.500 mbar.
The value displayed will be
GREEN
if the vacuum is ≤ 1.500 mbar.
Line 3: Vacuum Set Point
Underneath the vacuum measurement the current vacuum set point in mbar, Pa, or
Torr, as selected in Settings.
Line 4: Operation Mode
Indicates the current status of the freeze dryer. One of the following will always be
displayed on this line:
1.
Standby
– Collector and Vacuum are both OFF.
2.
Manual
– Collector and/or Vacuum are ON, but no Program is currently running.
3.
Program
– If a program is currently running this line will display the actual
Program Name, Step #, and Time Remaining (hh:mm) for the current step of the
program.