Labconco Glassware Washer Service/Technical Manual
18
6.
Replace the back paneling.
7.
Attach the hose from the valve to a Twin Inlet High Flow Air-gap
suitable for dishwasher applications. This may be obtained from a
plumbing supply store.
8.
Attach the drain hose from the Glassware Washer to the other inlet of
the Air-gap.
9.
Attach the Air-gap to the building drain. And turn back on the
electrical power.
Installation of the Optional Liquid Detergent Dispenser
An optional liquid detergent dispenser is available for installation on
SteamScrubbers® and FlaskScrubbers®. This permits dispensing liquid
detergent from a bulk 1 gallon (3.8L) container eliminating the need to fill
the detergent cup with detergent each time the washer is used.
To install the Liquid Detergent Dispenser:
1.
Disconnect power to the washer.
2.
Remove the back panel. Loosen the insulation jacket to provide
access to the back of the washer.
3.
Attach the dispense pump to the rear flange of the base using the
screws and nuts provided.
4.
Attach the tube from the detergent container to the inlet of the
pump.
5.
Loosen the hose clamp that is around the large hose on the bottom of
the vacuum break. DO NOT REMOVE THE HOSE. Slip the stainless steel
tube that is on the end of the clear tube coming from the detergent
dispense pump into the hose on the vacuum break. Tighten the hose
clamp. Secure the provided tie wrap around all 3 tubes on the top
of the vacuum break being careful not to collapse any tube.
6.
Locate the detergent container at a convenient location where it
will not be a trip hazard.
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com