― 34 ―
<Menu setting 1: SEt.1 item>
1) "Recording mode
Select one of the four recording mode.
[LOG] : Continuous recording mode
Make measurements and record the data at the selected
intervals.
[dtc] : Event recording mode
Can record total 8data(approx. for 0.8sec.): prior to and
subsequent to when exceeding the set current value.
The
LED for the CH on which the exceeding value is detected will
blink.
[̲ ̲̄] : Max. value recording mode
When exceeding the set value, record the measured current
value at every 10sec. The recording will be continued until the
measured value drops to 50% or less of the set value or for
10min. LED for the detected cannel will blink.
[CAP] Capture recording mode
Can record the instantaneous value when exceeding set
current value. Capable of observing waveform by transferring
the data to PC.
2-1) "Recording interval"
Recording interval can be selected in Continuous recording
mode.
Recording interval can be selected from:
1, 2, 5, 10, 15, 20, 30sec.
1, 2, 5, 10, 15, 20, 30, 60min.
2-2) "Current value for detection"
In Event, Max value and Capture recording mode, set the
current value for detection for each channel. When detecting
the current which exceeds this set value, record will start. Aldo,
LED blinks.
3) "One-time switch"
On : Recording will stop when memory is full.
Off : Overwrite the old data, and store the latest data (Endless)