― 23 ―
(8) Then, present status of one-time system is displayed. In case of
change, follow the procedure described in clause 9-2.Change of
set value.
One-time system can be selected either:
On : Recording will stop when memory is full.
Off : Overwrite the old data,and store the latest data
(9) By pressing "RANGE/▼"button, setting 1 is complete and "End"
message is displayed. Can back to "SEt.1" at first setting by
pressing "MENU/ENTER" or "START/STOP, CANCEL" button.
(10)Can back to initial measurement mode by pressing
"START/STOP, CANCEL" button.
Step3 : Clamp sensor connection, Preparation for measurement
(11)Connect the instrument and clamp sensor without clamping on
anything. Pay attention to the direction of connector and firmly
connected to the instrument.In capture recording mode, only Ch1
is used.
(12)Clamp the measured object on clamp sensor.
(13)Confirm that a sensor is connected to the instrument. When a
sensor is not connected, "nc"(non-connect) is displayed on the
LCD. (The measured value will not be recorded when sensor is
not connected at starting to record.)