― 15 ―
6-2 Event recording mode: Record with the set current of 15mA
Step1 : Power on
(1) Press "MENU/ENTER" button for 1sec. or more and power on the
instrument.
(2) Release the button when all indications are displayed on the LCD.
(3) When instrument is powered on, it becomes measurement mode.
Step2 : Check and change of set value
(4) Press "MENU/ENTER" button and make the instrument to menu
mode.
(5) Check Setting1:"SEt.1" is displayed on the LCD, and press
"MENU/ENTER" button.
(6) Present recording mode is displayed on the LCD. Check "dtc" is
displayed on the LCD, and press "RANGE/▼"button.
(When "LOG", "̲ ̲̄" and "CAP" is displayed on the LCD,
change the recording mode according to the procedure described
in clause 9-2.Change of set value.
Recording mode can be selected from:
LOG : Continuous recording mode
dtc : Event recording mode
̲ ̲̄ : Max.value recording mode
CAP : Capture recording mode
(7) Present current set value on CH1 is displayed on the LCD.
Confirm that the set current is 15mA, and press "RANGE/▼
"button.
In case of change, follow the procedure described in clause
9-2.Change of set value.
The current value can be set at every 1mA from 0 to 1000mA.
When the first current is detected, next detection will not start
until the detected value drops to 50% or less of set value. So
set the value that is more appropriate than that of measured
before stating to record.
(8) Check/change the current set value on CH2 and CH3 as well.