5.2.4.2.2 Shading
Reading Sensors on the scanner can be calibrated correctly with “Shading”.
Press the
R1
Key to highlight “Execute”, and then press the
Start
Key to start.
The following message is displayed.
Set the Calibration Sheet to the scanner, and then press the
Enter
Key to start.
Be careful of the direction of Calibration Sheet.
A corner not right angled must be on left rear against the direction to insert.
This corner mist be on left rear.
Direction to insert
Calibration Sheet