17
QuickGene-Mini8L Operation Manual
2 Operation
(5) Set the Cartridges in the Cartridge Holder so that the
position (No.) of each tubes is the corresponding
position in (3). Check the location of Waste tubes, 1.5
ml microtubes and Cartridges.
Cartridge Holder
Cartridge
Important
:
・
Align the taps on the Cartridge with the grooves
into the notch of Cartridge Holder and then push into
the bottom position. If the Cartridge is not inserted in
correct position, it may cause the pressure leakage.
・
If their positon does not correspond, Buffer and/or
lysate may spill. This results in an extraction failure,
and Buffer and/or lysate becoming waste.
Make sure they are
in relative position.
microtubes
No.5 No.6 No.7 No.8
No.5 No.6 No.7 No.8
No.1 No.2 No.3 No.4
No.1 No.2 No.3 No.4
Cartridge
Waste Tube
【
Cartridge Holder
】
(
Upper Row
)
【
Tube Holder
】
(
Lower Row
)
2
No.5 No.6 No.7 No.8
No.1 No.2 No.3 No.4
Summary of Contents for ADS Biotec QuickGene-Mini8L
Page 2: ......
Page 24: ...22 QuickGene Mini8L Operation Manual ...
Page 27: ...25 QuickGene Mini8L Operation Manual ...
Page 29: ...QuickGene Mini8L Operation Manual 1 Setting Up xxvi ...
Page 30: ......
Page 37: ...7 QuickGene Mini8L Operation Manual 1 Setting Up Plug AC Adapter ...
Page 44: ...12 QuickGene Mini8L Operation Manual 2 Operation Plug ...
Page 53: ......
Page 60: ...26 QuickGene Mini8L Operation Manual 2 Operation safety manuals ...
Page 61: ...27 QuickGene Mini8L Operation Manual 3 Maintenance 2 ...
Page 65: ......
Page 68: ...32 QuickGene Mini8L Operation Manual 4 Troubleshooting 4 ...
Page 69: ...33 QuickGene Mini8L Operation Manual 4 Troubleshooting 4 ...
Page 71: ...35 QuickGene Mini8L Operation Manual 5 Main Specifications 5 ...
Page 72: ......
Page 76: ......