21
QuickGene-Mini8L Operation Manual
2 Operation
(12) Repeat the wash process the necessary number of
times in correct rows according to the protocols.
Note
:
Refer to Step (10) to (11).
(13) After wash process is completed, return the rotary
switch to the
“OFF” position to raise the pressure
heads.
2
Important
:
·
Use both hands to rotate the rotary switch.
·
Make sure that no liquid remains in the
Cartridges.
·
Make sure that the Pressuring time of the last
wash process for at least 70 seconds. If the
Pressuring time was changed, return to original
setting (70 seconds) through the Pressure Mode
Switch (Refer to page 7 [Pressure Control
Switch] and 9 [Pressure Parameters]).
Note
:
Do not pull out the Holder unit while the pressure
heads are lowered.
(14) Pull out the Cartridge/Tube Holder.
・
Summary of Contents for ADS Biotec QuickGene-Mini8L
Page 2: ......
Page 24: ...22 QuickGene Mini8L Operation Manual ...
Page 27: ...25 QuickGene Mini8L Operation Manual ...
Page 29: ...QuickGene Mini8L Operation Manual 1 Setting Up xxvi ...
Page 30: ......
Page 37: ...7 QuickGene Mini8L Operation Manual 1 Setting Up Plug AC Adapter ...
Page 44: ...12 QuickGene Mini8L Operation Manual 2 Operation Plug ...
Page 53: ......
Page 60: ...26 QuickGene Mini8L Operation Manual 2 Operation safety manuals ...
Page 61: ...27 QuickGene Mini8L Operation Manual 3 Maintenance 2 ...
Page 65: ......
Page 68: ...32 QuickGene Mini8L Operation Manual 4 Troubleshooting 4 ...
Page 69: ...33 QuickGene Mini8L Operation Manual 4 Troubleshooting 4 ...
Page 71: ...35 QuickGene Mini8L Operation Manual 5 Main Specifications 5 ...
Page 72: ......
Page 76: ......