Installation and putting into operation of the BNA…EX bypass 
level indicator, heated, double-walled version, in explosion risk 
areas 

The  bypass  tube  is  also  available  in  a  heated,  double-walled  version,  in  which  the 

bypass  tube  is  surrounded  by  a  second  tube  (heating  jacket).  A  heat-carrying  medium,  heated 
liquid  or  heated  steam  can  be  pumped  through  the  resulting  double-mantle  space  through  two 
connections. Stainless material qualities in accordance with the AD specification sheets are used 
for the higher temperatures to which this construction is exposed. 

 

 

To  eliminate  the  risk  of  explosion  caused  by  the  heating  system,  the  following 
requirements apply to operation of the heating system and the heat-carrying medium: 

 

1.  The temperature of the heat-carrying medium must not exceed 80% of the ignition temperature 

of the stored material. 

2.  The equipment operator must ensure, by means of constant monitoring and operational checks, 

that  the  ignition  temperature  listed  under  item  1  is  not  exceeded.  These  monitoring  and 
checking activities must also take temperature dynamics resulting from chemical reactions into 
consideration. For example, maintenance of the permissible temperature level can be ensured 
on the basis of saturated steam pressure or the boiling point of a liquid. 

 

 

Summary of Contents for BNA EX Series

Page 1: ...Rev 7 16 03 2015 Manual de Instru es e de Montagem KEMA 02ATEX2106 X Por favor guarde esta informa o para uso futuro Please retain for future usage BNA EX Montage und Betriebsanleitung Mounting and op...

Page 2: ......

Page 3: ...OI_BNA_EX_pt_en doc...

Page 4: ...OI_BNA_EX_pt_en doc...

Page 5: ...OI_BNA_EX_pt_en doc...

Page 6: ...OI_BNA_EX_pt_en doc...

Page 7: ...OI_BNA_EX_pt_en doc...

Page 8: ...ratura 11 Dados de press o 11 English 12 Symbol legend 12 Safety information 12 Danger 13 Application and field of use 13 Attention 13 Removal of transport packaging and transport safety devices 13 In...

Page 9: ...vor leia estas Instru es antes de instalar e utilizar indicadores de n vel tipo Bypass BNA EX Este Manual destina se a profissionais que realizam a montagem a instala o e a configura o As normas de se...

Page 10: ...rui o dos indicadores de n vel tipo Bypass e causar danos pessais e materiais Todos os materiais do tubo Bypass e do flutuador devem ser resistentes ao meio a ser monitorado Os valores m ximos indicad...

Page 11: ...favor consulte os valores de torque dos parafusos Deve se aplicar selagens apropriadas Verificar se o material de selagem contra o meio e os seus vapores a temperatura e cargas de press o previstas s...

Page 12: ...segundo os folhetos AD Para eliminar o risco de explos o devido ao aquecimento aplicam se os seguintes requisitos para a opera o de aquecimento e para a transfer ncia de calor 1 A temperatura de trans...

Page 13: ...transdutores MG A descri o que se segue pretende ser apenas um guia para uma op o de montagem em determinado local Por favor leia tamb m as instru es de instala o e funcionamento de equipamento desmo...

Page 14: ...mplo de montagem MNAV montagem do interruptor magn tico em barra de reten o A montagem deste interruptor magn tico feita em uma barra de reten o separada 1 Retirar a barra de reten o 1 desapertando os...

Page 15: ...l tipo Bypass s pode ser garantida com a utiliza o de acess rios originais KSR Kuebler e pe as de reposi o Solu o de problemas Na tabela a seguir est o listados os problemas mais comuns e as contramed...

Page 16: ...cos a o inoxid vel 180 C temperatura m dia SG com escala alum nio gravado VSG com escala a o inoxid vel gravado especifica o somente se a op o estiver dispon vel S se pode instalar interruptores magn...

Page 17: ...9 21 OI_BNA_EX_pt_en doc extremidade da solda S S 3 4 Tipo b sico BNA...

Page 18: ...sico Flutuador cilindro Execu o sem gr nulos Tipo b sico Code 1 Code 2 Code 3 Code 4 Code 5 Code 6 Code 7 Z V S 250 16 60 1000 Code 7 Sistema magn tico Code 6 Densidade em kg m Code 5 Temperatura em...

Page 19: ...m xima Temperatura ambiente m xima no dispositivo de avalia o MRA MRAN MNAV MRK MNKV T1 320 C 320 C 50 80 C A temperatura ambiente m xima do dispositivo de avalia o deve ser respeitada A mesma n o de...

Page 20: ...G or MG Safety information Read these instructions before installing the BNA EX and putting them into operation These instructions are intended for the specialists in charge of mounting installation a...

Page 21: ...vels may result in malfunction or destruction of the bypass level indicator as well as to personal injury and property damage All of the materials used in the bypass chamber and float must be resistan...

Page 22: ...nce with any Special Conditions stipulated in such laws or regulations is of particular importance Please comply with the maximum torque ratings of the bolts screws used when tightening them down Use...

Page 23: ...ch this construction is exposed To eliminate the risk of explosion caused by the heating system the following requirements apply to operation of the heating system and the heat carrying medium 1 The t...

Page 24: ...evel Sensor MG The description provided here must be considered as being a means of orientation for the mounting on possibility provided by location specific conditions Please also refer to the assemb...

Page 25: ...witch must then be mounted the other way around with the type plate upside down Example of installation MNAV mounting of magnetic switch on a retainer rod This magnetic switch is attached to a retaine...

Page 26: ...level indicator can only be guaranteed if original KSR Kuebler accessories and replacement parts are used Error Search The following table lists the most frequent causes of error and the necessary cou...

Page 27: ...ature MNKV Magnetic Roller Indicator Edelstahl 180 C ambient temperature SG with Scale Aluminium engraved VSG with Scale Stainless steel engraved Only specification if option is available Only BGUV ma...

Page 28: ...Titanium max 16 bar Basic type Cylindrical floats Design without beads Basic type Code 1 Code 2 Code 3 Code 4 Code 5 Code 6 Code 7 Z V S 250 16 60 1000 Code 7 magnet system Code 6 density in kg m Code...

Page 29: ...processing and display equipment MRA MRAN MNAV MRK MNKV T1 320 C 320 C 50 80 C Avoid temperatures in excess of the maximum ambient temperature allowed for the installed processing and display equipmen...

Page 30: ..._en doc 22 21 KSR KUEBLER AG Adressen KSR KUEBLER Niveau Messtechnik AG Heinrich Kuebler Platz 1 D 69439 Zwingenberg Neckar Tel 49 06263 870 Fax 49 06263 87 99 http www ksr kuebler com e Mail info ksr...

Reviews: