7.3
Stopping operation
-
Press push button [1]
-
All LEDs are off and the device stops all functions
-
Set the toggle switch at the rear to "0" (off)
-
Close the hose clip of the refilling system
-
If not in use (for more than 60 min.) Disconnect the mains plug
8
Faults, troubleshooting
8.1
General
The ultrasonic nebulizer underwent a thorough quality check at the factory. If a fault occurs despite these tests, you
may be able to rectify them yourself if you observe the following notices.
There is a fault indicator on the left side of the control panel at the ON/OFF switch. This lights up red if a fault occurs
or the set time has elapsed (timer end).
After completing the set time see5.6
After a fault occurred, switch the device off.
Press the ON/OFF switch to do this. Remove the cause of the fault if possible and switch the device back on. The
following can be potential causes of the fault:
Fault
Cause
Remedy
Red light is lit
Not possible to switch on the
device
The plug is not positioned correctly in
the device or power socket
Check position of the plug in the
device or power socket
On/off switch at the rear of the device
is not set to On (I)
Set on/off switch to On (I)
On/off switch is switched off
Set switch to On
No, defective or wrong mains fuse
Check main fuse
Replace fuse as described in 8.3
Electronics faulty
Have it repaired
Quartz module with thermal
shutdown
Thermal switch has triggered. The
device is functional again after
cooling down for 10 minutes
Device is not working/
Red light is lit
Nebulizer chamber not fitted correctly Insert nebulizer chamber correctly.
Observe the preferred position
Nebulizer chamber empty
Refill nebulizer fluid
Leak in nebulizer chamber
Screw quartz oscillator firmly into
the nebulizer chamber
Faulty quartz oscillator
Replacing the quartz oscillator
Quartz module with thermal
shutdown
Thermal switch has triggered. The
device is functional again after
cooling down for 10 minutes
Electronics faulty
Have it repaired
Not enough nebulization when set
at maximum output
Nebulizer fluid does not match scale
Refill nebulizer fluid or remove some
Resistance of filter too high
Replace filter
Condensate in hose
Drain condensate. Re-route hose in
such a way that condensate can flow
back into the nebulizer chamber.