Operation
5.1 Configuration
For configuration connect the sensor to the loop and configure the sensor via PACTware
TM
FDT/DTM. Otherwise for offline configuration the sensor can be connected to the
OPTIBRIDGE/SMARTBRIDGE or another suitable USB interface cable.
OPTIBRIDGE/SMARTBRIDGE will be connected to the PC via USB port. Add the devices regarding
the instructions of PACTware
TM
.
For more information refer to
Overview Basic-DD menu tree (for Field Communicator 475)
on
page 39.
5.2 Calibration
For offline calibration via PACTware
TM
FDT/DTM the sensor can be connected to the
OPTIBRIDGE/SMARTBRIDGE or another suitable USB interface cable.
OPTIBRIDGE/SMARTBRIDGE will be connected to the PC via USB port. Add the devices regarding
the instructions of PACTware
TM
.
For more information refer to
Basic-DD menu tree (details for settings)
on page 40.
INFORMATION!
The sensor configuration has to be done in the online configuration dialog of PACTware
TM
FDT/DTM. If the configuration will be done in the offline configuration dialog of PACTware
TM
,
many functions are disabled e.g. calibration. The sensor configuration data will also be
overwritten with default settings for e.g. 4...20 mA, TAG, etc.
INFORMATION!
Set temperature compensation before calibration. Select between linear, natural water or off.
Default setting: linear