5
OPERATION
26
SMARTPAT COND 1200
www.krohne.com
09/2015 - 4003041501 - MA SMARTPAT COND1200 R01 en
i
The old cell constant and the calculated new cell constant appear
•
If you want to save the new cell constant press OK
OK
OK
OK to proceed
•
Set the calibration date with MM-DD-YYYY
•
Press Enter
Enter
Enter
Enter to proceed
i
The message Save value?
Save value?
Save value?
Save value? appears
•
Select Yes
Yes
Yes
Yes to save the values (Selecting No
No
No
No is followed by the same procedure just without
saving the values)
•
Press Enter
Enter
Enter
Enter to proceed
•
Deactivate the function Manual hold?
Manual hold?
Manual hold?
Manual hold? with choosing No
No
No
No
•
Press Enter
Enter
Enter
Enter to leave the calibration menu
Input cell constant
•
Choose Input cell constant
Input cell constant
Input cell constant
Input cell constant and press the key >
>>>
•
Activate the function Manual hold?
Manual hold?
Manual hold?
Manual hold? to avoid an alarm
•
Select Yes
Yes
Yes
Yes
•
Press Enter
Enter
Enter
Enter to proceed
•
Input the new cell constant
•
Press Enter
Enter
Enter
Enter to proceed
i
The message Save new cell constant?
Save new cell constant?
Save new cell constant?
Save new cell constant? appears and the values for the old and new cell
constant are shown
•
Press OK
OK
OK
OK to proceed
•
Set the calibration date with MM-DD-YYYY
•
Press Enter
Enter
Enter
Enter to proceed
i
The message Save value?
Save value?
Save value?
Save value? appears
•
Select Yes
Yes
Yes
Yes to save the values (Selecting No
No
No
No is followed by the same procedure just without
saving the values)
•
Press Enter
Enter
Enter
Enter to proceed
•
Deactivate the function Manual hold?
Manual hold?
Manual hold?
Manual hold? with choosing No
No
No
No
•
Press Enter
Enter
Enter
Enter to leave the calibration menu
CAUTION!
Moisture on the sensor connector must be avoided! Moisture may cause a short-circuit and a
malfuntion of the sensor!
If moisture has entered the connector dry it with air (e.g. hot air gun).
CAUTION!
•
Do not make any mechanical modifications to the sensor (electrodes shortened, drilled, bent
or scratched).
•
Make sure that the measuring electrodes are clean and dust-free. If necessary, clean the
electrodes as described on page 29
.
MA_SMARTPAT_COND_1200_en_150902_4003041501_R01.book Page 26 Monday, September 21, 2015 4:33 PM