31
6 Set up with the display and adjustment module
OPTISOUND 3010 C • Four-wire 4 … 20 mA/HART
30504-EN-220310
Up to 3000 measured values are recorded (depending on the sensor)
when starting a "
Trend curve
". Then the values can be displayed on
a time axis. The oldest measured values are always deleted.
In the menu item "
Choose curve
", the respective curve is selected.
Curve selection
Echo curve
►
Information:
The trend recording is not activated when being shipped. It must be
started by the user via the menu item "
Start trend curve
".
A comparison of the echo curve and the false echo curve allows a
more detailled evaluation of measurement reliability. The selected
curve is updated continuously. With the
[OK]
key, a submenu with
zoom functions is opened.
The following functions are available with "
Echo and false echo
curve
":
•
"X-Zoom": Zoom function for the meas. distance
•
"Y-Zoom": 1, 2, 5 and 10x signal magnification in "
dB
"
•
"Unzoom": Reset the presentation to the nominal measuring range
without magnification
In the menu item "
Trend curve
" the following are available:
•
"X-Zoom": Resolution
– 1 minute
– 1 hour
– 1 day
•
"Stop/Start": Interrupt a recording or start a new recording
•
"Unzoom": Reset the resolution to minutes
As default setting, the recording pattern has 1 minute. With the adjust-
ment software PACTware, this pattern can be also set to 1 hour or 1
day.
Echo curve
High nozzles or vessel installations, such as e. g. struts or agitators as
well as buildup and weld joints on the vessel walls, cause interfering
reflections which can impair the measurement. A false echo storage
detects and marks these false echoes, so that they are no longer
taken into account for the level measurement. A false echo memory
should be created with low level so that all potential interfering reflec-
tions will be detected.
Diagnosis - Curve pres-
entation
Service - False signal
suppression
Summary of Contents for OPTISOUND 3010 C
Page 1: ...OPTISOUND 3010 C Handbook Ultrasonic Level Transmitter Four wire 4 20 mA HART...
Page 49: ...49 Notes OPTISOUND 3010 C Four wire 4 20 mA HART 30504 EN 220310...
Page 50: ...50 Notes OPTISOUND 3010 C Four wire 4 20 mA HART 30504 EN 220310...
Page 51: ...51 Notes OPTISOUND 3010 C Four wire 4 20 mA HART 30504 EN 220310...