INSTALLATION
16
OPTISENS TUR 2000
www.krohne.com
07/2017 - 4001970202 - MA OPTISENS TUR 2000 R01 en
3.10 Mounting the sensor into an assembly
For further instructions on installation into an immersion assembly refer to the assembly
manual.
Use an assembly that does not fix the sensor cable or require the sensor to be screwed into the
assembly. If the sensor needs to be screwed for mounting or dismounting make sure that the
sensor cable is turned into the same direction.
Installing procedure
•
Insert the sensor cable through the immersion assembly.
•
Fasten the sensor to the tip of the telescopic rod by tighteninng the 6 screws of the 2 holding
plates.
•
Connect the wires either to the control system directly (only 4...20 mA) or to the MAC 100
signal converter. For further information refer to
Connecting the sensor to PC
on page 22,
refer to
Connecting the sensor directly to the control system
on page 22 and refer to
Connecting the sensor cable to the signal converter
on page 20.
For removing the sensor, repeat the steps above in reverse order.
Calibrate the sensor before installing it into the assembly. For further information refer to
Calibration
on page 26
DANGER!
All work on the electrical connections may only be carried out with the power disconnected.
CAUTION!
Do not turn the cable gland on the sensor this might cause a sensor leak and damage the
electronics inside. While mounting or dismounting the sensor, the senor cable must not be fixed
or trapped as this might loosen the water tight gland connection from the sensor.
INFORMATION!
Recalibrate the sensor after each manual cleaning procedure.