OPERATION
5
27
OPTISENS TUR 2000
www.krohne.com
07/2017 - 4001970202 - MA OPTISENS TUR 2000 R01 en
Step 1: Activating the hold function
•
You have activated the manual hold
manual hold
manual hold
manual hold function. Go to the next step and prepare the calibration
procedure. You have to return to the measuring mode.
•
Press
^
^
^
^
until you reach the measuring mode again.
Step 2: Preparing the calibration procedure
•
For re-calibration, remove the sensor from the process.
•
If you calibrate a sensor, make sure that the sensor is correctly connected to the signal
converter.
•
Check the sensor for damages, check the measurement windows for coating and rinse the
sensor tip with tap water and gently swipe it with a soft tissue.
•
Do not touch or scratch the windows.
•
Repeat the cleaning step twice to make sure that the sensor is clean as well from deposits we
can not detect with human eye (e.g. polymer film).
•
Press >>>> for more than 2.5 seconds, then release the button. You are on the main menu level.
•
Press
or
until the main menu quick setup
quick setup
quick setup
quick setup is highlighted.
MAIN MENU
> A quick setup
> A quick setup
> A quick setup
> A quick setup
B test
C setup
D service
•
Press >>>> to enter the chosen menu.
You are on the first submenu level. In the upper line of the display quick setup
quick setup
quick setup
quick setup appears, beneath the
submenu language
language
language
language is highlighted.
•
Press
or
until the submenu manual hold
manual hold
manual hold
manual hold is highlighted.
•
Press >>>> to enter the chosen menu.
You are on the second submenu level. In the upper line of the display manual hold
manual hold
manual hold
manual hold appears, beneath
the option off
off
off
off is highlighted.
•
Press
or
to choose the option on
on
on
on.
•
Press
^
^
^
^
to confirm the entered value.
INFORMATION!
Use a soft tissue to wipe the window clean. Otherwise the polymer films are not being removed
and interfere with later calibration.
INFORMATION!
Before proceding with product cal.
product cal.
product cal.
product cal. make sure that the offset
offset
offset
offset is set to 0.000 NTU