Step 3a: Accessing the calibration menu via the main menu setup
setup
setup
setup
Step 3b: accessing the calibration menu via the main menu quick setup
quick setup
quick setup
quick setup
•
Tape three to four layers of new black electrical isolation tape across the detector window.
•
While taping you will notice that the measuring value decreases.
•
Make sure that the detector is absolutely sealed from light sources.
•
Press >>>> for more than 2.5 seconds, then release the button. You are on the main menu level.
•
Press
or
until the main menu setup
setup
setup
setup is highlighted.
MAIN MENU
A quick setup
B test
> C setup
> C setup
> C setup
> C setup
D service
•
Press >>>> to enter the chosen menu.
You are on the first submenu level. In the upper line of the display setup
setup
setup
setup appear. Press
or
until the
submenu process input A
process input A
process input A
process input A is highlighted.
•
Press >>>> to enter the chosen menu.
You are on the second submenu level. In the upper line of the display process input A
process input A
process input A
process input A appear. Press
or
until the submenu product cal.
product cal.
product cal.
product cal. is highlighted.
•
Press >>>> to enter the chosen menu.
The actual measured value is displayed.
•
Press >>>> for more than 2.5 seconds, then release the button. You are on the main menu level.
•
Press
or
until the main menu quick setup
quick setup
quick setup
quick setup is highlighted.
MAIN MENU
> A quick setup
> A quick setup
> A quick setup
> A quick setup
B test
C setup
D service
•
Press >>>> to enter the chosen menu.
You are on the first submenu level. In the upper line of the display quick setup
quick setup
quick setup
quick setup appear. Press
or
until
the submenu product cal.
product cal.
product cal.
product cal. is highlighted.
•
Press >>>> to enter the chosen menu.
The actual measured value is displayed.
Step 4: Preparing the zero point calibration
Figure 5-1: Preparing the zero calibration
1
Detector
2
Ermitter
3
Black electrical isolation tape