Step 4a: accessing the calibration menu via the main menu setup
setup
setup
setup
•
Press >>>> for more than 2.5 seconds, then release the button. You are on the main menu level. In the upper line of the
display "A" appears, beneath the main menu quick setup
quick setup
quick setup
quick setup is highlighted.
•
Press
or
until the main menu setup
setup
setup
setup is highlighted.
MAIN MENU
A quick setup
B test
> C setup
> C setup
> C setup
> C setup
D service
•
Press >>>> to enter the chosen menu.
You are on the first submenu level. In the upper line of the display "setup" and "C1" appear. Press
or
until the submenu process input A
process input A
process input A
process input A is highlighted.
•
Press >>>> to enter the chosen menu.
You are on the second submenu level. In the upper line of the display "process input A" and "C1.1"
appear. Press
or
until the submenu product cal.
product cal.
product cal.
product cal. is highlighted.
•
Press >>>> to enter the chosen menu.
You are on the third submenu level. In the upper line of the display "product cal." and "C1.2"
appear.
The submenu air pressure
air pressure
air pressure
air pressure is highlighted.
•
Press
or
to select the air pressure in Pa.
•
Press
^
^
^
^
to confirm.
The submenu humidity
humidity
humidity
humidity is highlighted.
•
Press
or
to select the humidity in %.
•
Press
^
^
^
^
to confirm.
The submenu salinity
salinity
salinity
salinity is highlighted.
•
Press
or
to select the salinity in mg/L.
•
Press
^
^
^
^
to confirm.
The submenu sensitivity calibration
sensitivity calibration
sensitivity calibration
sensitivity calibration is highlighted.
•
Press
or
until the submenu start the calibration
start the calibration
start the calibration
start the calibration is highlighted.
•
Press
^
^
^
^
to start the calibration.
The actual oxygen is displayed. Please wait till the value is stable.
•
Press
^
^
^
^
to proceed.
There is a count down of 35 seconds for the sensitivity calibration.
The message sensitivity calibration successful
sensitivity calibration successful
sensitivity calibration successful
sensitivity calibration successful is highlighted if the calibration was successful.
•
Press
^
^
^
^
to proceed with the zero point calibration.
The message sensitivity calibration unsuccessful
sensitivity calibration unsuccessful
sensitivity calibration unsuccessful
sensitivity calibration unsuccessful is highlighted if the calibration was not
successful.
•
Press
or
to choose between repeat the calibration yes or no.
The submenu zero point calibration
zero point calibration
zero point calibration
zero point calibration is highlighted.
•
Press
^
^
^
^
to start the zero point calibration.
The actual concentration is displayed. Please wait till the value is stable.
•
Press
^
^
^
^
to confirm.
The message zero point calibration successful
zero point calibration successful
zero point calibration successful
zero point calibration successful is highlighted if the calibration was successful.
•
In this case press
^
^
^
^
several times to leave the calibration menu.
The message zero point calibration unsuccessful
zero point calibration unsuccessful
zero point calibration unsuccessful
zero point calibration unsuccessful is highlighted if the calibration was
unsuccessful.