INSTALLATION
3
15
OPTISENS ODO 2000
www.krohne.com
10/2016 - 4003689502 - MA OPTISENS ODO 2000 R02 en
3.9 Sensor use with cleaning function
When the automatic cleaning is required the device can be equipped with a cleaning hose for air
cleaning.
Before the device is installed into its final measuring location the following points must
be observed
•
provide hose (Ø 6.3...9.5 mm / 1/4...3/8") in suitable length
•
prepare a connection for the cleaning hose
•
push the hose onto the cleaning connector
•
put the sensor cable and cleaning hose through the adequate extension pipe of the mounting
assembly
The pressurised air is to be provided and must be clean with a max of 3 bar.
The typical cleaning time is 15 seconds and the typical cleaning frequency is 2 times/day, but this
will differ from application to application.
INFORMATION!
Most applications do not require the cleaning function if the sensor is installed correctly in a
correct angle.