OPERATION
5
37
OPTISENS ODO 2000
www.krohne.com
10/2016 - 4003689502 - MA OPTISENS ODO 2000 R02 en
5.3.3 Digital mode
Command Format: ID + M + x
^
For example, if ID = 14, and analog out = ppm type 14M1
^
or 00M1
^
The probe can be configured for digital mode (digital mode=1) or analog 4...20 mA (digital
mode=0).
5.3.4 Analog output
Command Format: ID + O + x
^
For example, if ID = 14, and analog out = ppm type 14O1
^
or 00O1
^
The analog output 4...20 mA can be linked to the %sat or ppm scale:
Set the parameter x=0 for ppm
x=1 for the %sat scale
5.3.5 Chloride salinity
Command Format: ID + C + x
^
For example, if ID = 14, and salinity 20000 ppm type 14C200
^
or 00C200
^
To check whether the entered value was received, type the command 'A'.
5.3.6 Atmospheric pressure
Command Format: ID + P + x
^
For example, if ID = 14, and the pressure 780mmHg, type 14P780
^
or 00P780
^
To check whether the entered value was received, type the command 'A'.
5.3.7 Relative humidity
Command Format: ID + U + x
^
For example, if ID = 14, and humidity 50%, type 14U50
^
or 00U50
^
To check whether the entered value was received, type the command 'A'.
Probe response:
ID + C + x
command executed correctly
Probe response:
none
command does not run properly
Probe response:
ID + U + x
command executed correctly
Probe response:
none
command does not run properly
Probe response:
ID + U + x
command executed correctly
Probe response:
none
command does not run properly