5.3.8 Sensitivity calibration
Command Format: ID + S
^
For example, if ID = 14, type 14S
^
or 00S
^
To check whether the entered value was received, type the command 'A'.
5.3.9 Zero point calibration
Command Format: ID + Z
^
For example, if ID = 14, type 14Z
^
or 00Z
^
To check whether the entered value was received, type the command 'A'.
5.3.10 Last calibration date
Command Format: ID + D + cccccccc
^
For example, if ID = 14, and we need to set the date 13/11/10, type 14D13/11/10
^
or 00D13/11/10
^
The Command is for memorising the last calibration date.
The operator can use the 8 characters as he likes better (no syntaxes limitation).
5.3.11 ID value
Command Format: ID + I + x
^
For example, if ID = 14, and we need to set the ID at 07, type 14I07
^
or 00I07
^
The new ID will be available from the next start up of the probe.
Probe response:
ID + S + x
command executed correctly
Probe response:
none
command does not run properly
Probe response:
ID + Z + x
command executed correctly
Probe response:
none
command does not run properly
Probe response:
ID + D + cccccccc
command executed correctly
Probe response:
none
command does not run properly
Probe response:
ID + I + x
command executed correctly
Probe response:
none
command does not run properly