.52+1(
BM 70 A Kurzanleitung
Seite: 7
4. Entwurf 03/97
Signal output variants
Variant
Comments
Ex-e current output HART
®
Ex-e; with switching output and digital input
Ex-i current output HART
®
intrinsically safe; passive
Current output
Ex-e; not communicable
RS 485 + current output
BM70 A protocol or ModBus-RTU; Ex-e
PROFIBUS FMS/DP
RS485 interface; Ex-e; see Add’l Instructions
PROFIBUS PA
intrinsically safe; see Add’l Instructions
Terminal assignment
Supply power:
N
L
PE
~
~
N
L
Ein- und Ausgänge
1
2
+(-)
-(+)
24V DC
24V AC
FE
~
~
Ein- und Ausgänge
Connections for 115/230 V AC
Connections for 24 V DC/AC (any
polarity)
Connection of a functional ground FE is not
mandatory.
Current output HART
®
, Ex e:
Ex-i current output HART
®
:
32 31
-
+
42 41
Strom-
ausgang
Schaltausgang
max. 100mA/30V DCAC
Digital-
Eingang
82 81
- +
- +
5...28V=
4-20mA
max.500
Ω
32 31
-
Ex-i-Speise-
+
+
für externe
Einheit
+
-
U max.= 30V
Ex-i-Kreise
Strom-
ausgang
4-20mA
U min.= 8V
The
4 other terminals can be used for looping-through
external intrinsically safe circuits.
Current output
QRQFRPPXQLFDEOH
:
RS485 version (Ex-e):
32 31
Strom-
ausgang
- +
nicht belegte
Anschlüsse
-
+
4-20mA
max.250
Ω
B
B
A
A
RS 485
Eingang
RS 485
Ausgang
32 31
Strom-
ausgang
- +
-
+
4-20mA
max.250
Ω