Error display during measurement
When function 3.2.5 "ERROR. MSG." is set to YES, any error occurring during
measurement is indicated in the display and alternates with the measured value for
as long as the error is present.
In addition, all errors are stored. Press the keystroke combination
to get into the error list. You can page through the list with key
, and
acknowledge the errors at the end - if required - by "QUIT YES". Press key
twice
to return to the measuring mode.
Fatal errors (FATAL ERROR), that are detected when the device is started up,
render operation of the BM 70 A impossible.
6. Technical data (extract)
Tank height (measuring range)
0.5 to 40 m (Wave-Stick up to 20 m); larger on request
Measuring accuracy (distance)
Standard: from 1m: + 1 cm / from 3.3m: + 0.3%
BM 70 Precision: from 1m: ± 5 mm / from 5m: ± 0.1%
Measured-value resolution
1 mm
Rate of change in level
max. 10 m/min (tracking speed)
Connecting flanges
Horn antenna/Wave-Guide
DIN 2501 DN 50 to DN 200 / PN 6 to PN 64;
Shape C to DIN 2526 or others
ANSI B16.5 2” to 8", Class 150 lbs or 300 lbs, RF
Wave-Stick
Only DN 50...150 or ANSI 2”...6”, dairy DIN11851
DN60/65/80, Triclamp 2/3/4”, SMS 51/63/76 mm
Max. allowable operating pressure -1 bar (vacuum) to max. 64 bar, depending on version and
flange pressure rating, information on special versions > 64 bar
supplied on request
Flange system with horn antenna or Wave-Guide:
Connection: nominal dia.
Flange rated pressure
PN 16
PN 25
PN 40
PN 64
DN mm
inches
bar
psig
bar
psig
bar
psig
bar
psig
80
3
16
232
---
---
40
580
64
928
100
4
16
232
---
---
38
551
55
797
150
6
16
232
---
---
34
493
47
681
200
8
16
232
25 bar
362
32
464
45
652
Wave-Stick:
max. 16 bar, temperature-dependent:
Temperatur T [°C]
Druck
-20
+150
+100
-1
+1
+16
[bar]
46 - 0.3·T[°C]
bar