The LCD indication is different from internal fixation.
7-32
LR indication of the LCD is wrong.
(LR indication of the LCD does not accord with the right and left position of the optical head base.)
Cause
Remedy
The right-and-left detection board is
dirty.
Clean a right-and-left detection board. The cleaning
method of a right-and-left detection board should look at
P.40 "the cleaning method of right-and-left detection."
The R/L selection sensor is dirty.
Clean an R/L selection sensor. The cleaning method of a
R/L selection sensor should look at P.40 "the cleaning
method of right-and-left detection."
Faulty R/L selection sensor
Replace the R/L selection sensor
Cause
Remedy
The right-and-left detection board is
dirty.
Clean a right-and-left detection board. The cleaning
method of a right-and-left detection board should look at
P.40 "the cleaning method of right-and-left detection."
The R/L selection sensor is dirty.
Clean an R/L selection sensor. The cleaning method of an
R/L selection sensor should look at P.40 "the cleaning
method of right-and-left detection."
Faulty R/L selection sensor
Replace the R/L selection sensor
Summary of Contents for K9L39F5BE
Page 4: ...GG 09004 Edition 4 0 3 Names and Functions of Components...
Page 5: ...GG 09004 Edition 4 0 4...
Page 6: ...GG 09004 Edition 4 0 5...
Page 9: ...GG 09004 Edition 4 0 8 2 4 BLOCK DIAGRAM...
Page 10: ...GG 09004 Edition 4 0 9 2 Wiring Diagram 2 1 K9L39F5...
Page 11: ...GG 09004 Edition 4 0 10 2 2 K9L39F5B...
Page 12: ...GG 09004 Edition 4 0 11 2 3 K9L39F7 F8 H8 F9 H9...
Page 13: ...GG 09004 Edition 4 0 12 2 4 K9L39F7B F8B H8B F9B H9B...