Record Method (Data storage method)
Set storage method for measured data by channel to inner/external memory.
Display and print method is Record Method.
Set value
Description
Instant(Instant value)
Saves measuring value by every record period (Log Speed)
Average(Average value) Saves averaged measuring value during record period (Log Speed).
Minimum(Min. value)
Saves min. measuring value during record period (Log Speed).
Maximum(Max. value)
Saves max. measuring value during record period (Log Speed).
Set range: Instant
↔
Average
↔
Minimum
↔
Maximum
Factory default: Instant
8.1.17
Filter Type (Input digital filter)
In some applications the fluctuating measuring input causes the display value to fluctuate. In this
case accurate display/record is disable. This function is able to make display value stable by
input digital filter.
Input digital filter uses moving average method (Moving Average Filter). It does not affet to
display period but display value may be different with input value.
Set range: None
↔
Moving
Factory default: None
Summary of Contents for krn100
Page 1: ... Copyright Reserved KONICS Co Ltd i USER MANUAL ...
Page 2: ...ii Copyright Reserved KONICS Co Ltd ...
Page 12: ...Table of Contents xii Copyright Reserved KONICS Co Ltd ...
Page 26: ...2 Specification 26 Copyright Reserved KONICS Co Ltd ...
Page 28: ...3 Dimensions 28 Copyright Reserved KONICS Co Ltd ...
Page 32: ...4 Connection 32 Copyright Reserved KONICS Co Ltd ...
Page 45: ...7 Operation Copyright Reserved KONICS Co Ltd 45 7 Operation ...
Page 52: ...7 Operation 52 Copyright Reserved KONICS Co Ltd 7 3 2 Parameter ...
Page 54: ...7 Operation 54 Copyright Reserved KONICS Co Ltd ...
Page 144: ...9 DAQMaster 144 Copyright Reserved KONICS Co Ltd ...
Page 150: ...10 Maintenance 150 Copyright Reserved KONICS Co Ltd ...
Page 153: ...11 Troubleshooting Copyright Reserved KONICS Co Ltd 153 ...
Page 154: ......