45
REGIUS MODEL 190 OPERATION MANUAL Ver.1.03 2004.12
5
6
Operating
Initializing
Initializing
ERROR [04904]
BAD LAMP CALL SVC
Network Error [24000]
MAC : XXXXXXXXXXXX
Network Error [24000]
IP : XXX.XXX.XXX.XXX
REGIUS MODEL 190
STATUS
No.
ERROR MESSAGE
REMEDY
The error with this error code will be displayed when both
erasing lamps are broken. Once the error occurs, the REGIUS
MODEL 190 cannot be operated.
When this error occurs while initializing, both MAC address and
IP address are displayed alternately on the LCD panel. In this
case, follow the procedure below.
1. Ensure that the ethernet cable is connected to the REGIUS
MODEL 190 properly. If the cable is not connected, connect it.
Then the REGIUS MODEL 190 will start up automatically.
2. Ensure that the JM is operating and the ethernet cable is
connected to it.
a) If the JM is not operating:
Start up JM after checking Ethernet cable connection:
REGIUS MODEL 190 will start up automatically. Finally,
restart the REGIUS Console.
b) If the ethernet cable is not connected to JM:
Connect it to the JM. At this point, if the JM is operating,
the REGIUS MODEL 190 will start up automatically.
Finally, restart the REGIUS Console.
Handling errors with error codes
There are two types of JM; One is built in the
CS-1/CS-2/CS-3, the other is the external.
If one of the erasing lamps is broken, the
message "READY" together with "BAD LAMP
CALL SVC" is displayed on the LCD panel. This
status, under which the erase time becomes
longer, will not effect normal operation of
REGIUS MODEL 190. As a precaution, contact
your Technical Representative.
Summary of Contents for 0676
Page 8: ...6 REGIUS MODEL 190 OPERATION MANUAL Ver 1 03 2004 12...
Page 26: ...24 REGIUS MODEL 190 OPERATION MANUAL Ver 1 03 2004 12...
Page 44: ...42 REGIUS MODEL 190 OPERATION MANUAL Ver 1 03 2004 12...
Page 55: ...53 REGIUS MODEL 190 OPERATION MANUAL Ver 1 03 2004 12...
Page 68: ...0676 55010 E 0412TE...