- 14 -
1. Sensor
Sensor head selection and configuration of the settings of any sensor.
Setting channel selection
[Sensor]
Select using
△
or
▽
.
CH Setting channel lights
CH1: Channel 1
CH2: Channel 2
CH3: Channel 3
CH4: Channel 4
ALL: Simultaneous setting
of channels 1 through 4
SEQU: Sequential setting
of channels 1 through 4
Any.S: Any sensor
PH-1: MSU-PH-EA (-101.3 to 0.0 kPa)
PH-2: MSU-PH-ER (-100.0 to 220.0 kPa)
PH-3: MSU-PH-EM (-100 to 1000 kPa)
FS-1: FS-R05 (-500 to 500 mL/min)
FS-2: FS-R3 (-3.00 to 3.00 L/min)
FS-3: FS-10 (0.00 to 10.00 L/min)
Selects a sensor head.
Returns to:
[1. SEn]
Returns to:
Returns to:
SW
■■■■
Sensor head selection
CH Setting channel lights
SW
■■■■
Setting channel selection
[Any Sensor]
CH Setting channel lights
CH1: Channel 1
CH2: Channel 2
CH3: Channel 3
CH4: Channel 4
ALL: Simultaneous setting
of channels 1 through 4
SEQU: Sequential setting
of channels 1 through 4
[Allowable input range]
-9999 to 9999
Initial Default: 1000
0: No decimal point
0.0: One decimal place
0.00: Two decimal places
0.000: Three decimal places
Configures display settings
of any sensor
(when SE-0 is selected).
SW
■■■■
Decimal point position setting
CH Setting channel lights
SW
□■■■
Display setting for 5V
CH Setting channel lights
SW
■□■■
[Allowable input range]
-9999 to 9999
Initial Default: -1000
Display setting for 1V
CH Setting channel lights
SW
■■□■
[Unit Segment Indicator
Lights]
None
kPa
L/min
mL/min
• Switches the segment next to the unit.
Display unit setting
CH Setting channel lights
SW
■■■□
Select using
△
or
▽
.
Setting channel selection
[Input CH]
CH Setting channel lights
CH1: Channel 1
CH2: Channel 2
CH3: Channel 3
CH4: Channel 4
ALL: Simultaneous setting
of channels 1 through 4
SEQU: Sequential setting
of channels 1 through 4
IN1: Sensor Channel 1
IN2: Sensor Channel 2
IN3: Sensor Channel 3
IN4: Sensor Channel 4
Initial factory defaults
CH1:IN1
CH2:IN2
CH3:IN3
CH4:IN4
Selects the sensor input
channel.
SW
■■■■
Sensor input channel selection
CH Setting channel lights
SW
■■■■
• After the sensor head setting is changed, the threshold value is
initialized to: Threshold Value = (Upper Limit Value + Lower
Limit Value)/2.
This means you will need to re-configure the
threshold value setting.
button
button
button
button
button
button
button
button
button
button
button
button
button
※※※
※※※
※
※
※