![Kobold NRM Operating Manual Download Page 15](http://html1.mh-extra.com/html/kobold/nrm/nrm_operating-manual_1982603015.webp)
NRM
NRM K06/0422
page 15
8.1.2
Information screens
Press button to cycle between the main
measurement screen and the information display
screen:
1. General information screen (DEV. INFO)
Overall running time (OV. RUN TIME)
Run time after power on (RUN TIME)
Type of interface (INTERFACE) in the instrument.
Type of the instrument (TYPE)
2. Sensor information screen: (SENSOR INFO)
Number of echoes (ECHO TOT/SEL)
Blocking (BLOCKING)
Signal-to-noise ratio (SN)
Temperature (TEMP)
3. Echo table: (ECHO TABLE)
The location (distance) and the amplitude of the
echoes (Dist. / Amp.) are listed
The listed items are the reflections detected by the
NRM (measured in dB) and the approximate
distance from the process connection. The listed
values are not accurate measurement values, since
around the selected echo (measurement window)
there are further measurements and signal
processing procedures in order to provide accurate
measurement display and level transmission.
The informative screen returns back to main screen after 30 seconds. By pressing
the button the user can return to the main screen any time. Pressing the
E
button in any of the screens the user can enter the main menu. After exiting the
menu always the main screen will be shown.
Summary of Contents for NRM
Page 1: ...Operating Instructions for Non Contact Radar Level Transmitter Model NRM...
Page 43: ...NRM NRM K06 0422 page 43 20 ATEX certificate...
Page 44: ...NRM page 44 NRM K06 0422...
Page 45: ...NRM NRM K06 0422 page 45...
Page 46: ...NRM page 46 NRM K06 0422...
Page 47: ...NRM NRM K06 0422 page 47...
Page 48: ...NRM page 48 NRM K06 0422...
Page 49: ...NRM NRM K06 0422 page 49...
Page 50: ...NRM page 50 NRM K06 0422 21 IECEx certificate...
Page 51: ...NRM NRM K06 0422 page 51...
Page 52: ...NRM page 52 NRM K06 0422...