Caution: Ensure the power is isolated to the circuit. Max. Input overload: 250V rms
<
10sec
1. Press “REL”, to measure the relative value: “REL” appears on the display, the auto range mode will
be changed to manual range mode. Press again, and “REL” disappears, you return to auto range
mode. NB: the meter cannot go back to auto range mode whilst in” Hz” measurement, and will not
measure the relative value
Testing Specifications
Accuracies are +/- (% of r number in last digit) at 23°C +/- 5°C
<
70% RH
DC Voltage
• Overload protection: 1000V DC/ 750V rms AC
• Impedance: 10 MΩ, more than 100 MΩ on 400mV range
AC Voltage
Range
Accuracy
Resolution
400mV
0.8% of rdg +5 digits
0.1mV
4V
1.0% of rdg +5 digits
1mV
40V
1.0% of rdg +5 digits
10mV
400V
1.0% of rdg +5 digits
100mV
1000V
1.5% of rdg +5 digits
1V
Range
Accuracy
Resolution
Frequency
4V
1.5% of rdg+5 digits
1mV
50~400Hz
40V
1.5% of rdg+5 digits
10mV
50~400Hz
400V
1.5% of rdg+5 digits
100mV
50~400Hz
750V
2.5% of rdg+5 digits
1V
50~100Hz
Range
Accuracy
Resolution
4V
3.0% of rdg+10 digits
0.1A
1000A 0~800
3.5% of rdg+10 digits
1A
1000A 800~1000
6.5% of rdg+10 digits
1A
• Average sensing, calibrated to rms of sinewave
• Overload protection: 1000V DC/750V rms AC
• Impedance: 10 MΩ
DC Current
• Overload protection 1000A rms within 60 seconds
RELATIVE MEASUREMENT: