4 Commissioning
WARNING! Process medium may leak from the SensoGate WA131M in the event of damage or
improper installation, and may contain hazardous substances.
Follow the safety instructions.
Note:
Upon request, Knick Elektronische Messgeräte GmbH & Co. KG will provide safety instruction
and product training during initial commissioning of the product. More information is available from
the relevant Knick representatives.
01.
Install the SensoGate WA131M.
Retractable Fitting: Installation, p. 22
02.
Install the outlet hose.
Outlet Hose: Installation, p. 23
03.
Optional: Install the inlet hose.
Inlet Hose: Installation, p. 23
04.
Mount the sensor.
Installing and Removing Sensors, p. 27
05.
Ensure that the process connection is securely fastened.
06.
Optional: Ensure that installed safety accessories (e.g., ZU0818 retainer clamp) are securely fas-
tened.
07.
Optional: Ensure that the SensoGate WA131M-X is correctly connected to the plant’s equipotential
bonding system.
Operation in Explosive Atmospheres, p. 9
08.
Move the SensoGate WA131M into the PROCESS position.
Moving into the PROCESS Position, p. 25
✓
Safety lock button pops out when the PROCESS position is reached.
✓
Rotating collar is locked to prevent rotation.
09.
Move the SensoGate WA131M into the SERVICE position.
Moving into the SERVICE Position, p. 26
✓
Safety lock button pops out when the SERVICE position is reached.
✓
Rotating collar is locked to prevent rotation.
10.
Check the SensoGate WA131M for leaks under process conditions.
✓
SensoGate WA131M and connections have no leaks.