SensoGate WA131M
28
07.
On first-time installa
tion: Hold the sensor cable
(6)
in a loop and fast
en it with the clamp
(7)
. Dur-
ing this process, the sensor cable loop must be long enough so that the sensor cable does not im-
pede the stroke movement of the SensoGate WA131M.
08.
On first-time installation: Optionally connect the equipotential bonding line
(8)
t
o the clamp
(3)
.
09.
Optional: Install ZU0759 protective cap.
5.3.3 Solid-Electrolyte Sensor, Short Immersion Depth: Removal
Note:
On versions with rinse connection, rinse the sensor prior to removal in order to prevent entrain-
ment of chemically aggressive process medium in the area of the sensor holders.
1
2
3
5
4
01.
Move the SensoGate WA131M into the SERVICE position.
Moving into the SERVICE Position, p. 26
02.
Check outlet and leakage bores for escaping process medium. If process medium escapes, depres-
surize the process and perform troubleshooting.
03.
Optional: Remove ZU0759 protective cap.
04.
Disconnect the cable bushing
(4)
of the sensor cable
(3)
fr
om the sensor head
(1)
.
05.
R
elease the sensor
(2)
using the spanning wrench
(5)
(
A/F 19 mm). R
ecommended tools: ZU0647
sensor spanning wrench
06.
P
ull out the sensor
(2)
.
07.
If the sensor glass is broken, check the immersion tube gasket for damage and replace it if
necessary
.