SensoGate WA131M
23
3.3 Outlet Hose: Installation
Note:
The outlet is used to drain trapped process medium and must not be closed. We recommend in-
stalling the supplied outlet hose even for versions without a rinse connection. By moving the sensor to
the respective limit positions, pressurized process medium can enter the calibration chamber and be
compressed when the outlet is closed. This process medium may splash out during sensor replace-
ment.
2
3
4
3
4
1
Note:
The symbol
(1)
designates the outlet.
01.
P
ush the hose clamp
(3)
ont
o the outlet hose
(4)
.
02.
Complet
ely push the outlet hose
(4)
onto the connec
tion nipple
(2)
.
03.
Secur
e the outlet hose
(4)
with the hose clamp
(3)
.
3.4 Inlet Hose: Installation
NOTICE!
Drinking water may be contaminated by rinse and process media when connecting to drink-
ing water pipes.
Observe the information con
tained in EN 1717
. Install a suitable check valve
(e.g
., check v
alv
e R
V01) at the water or rinse connection.
1
2
5
4
3
1
5
4
3
Note:
When using versions of SensoGate WA131M with inlet port, the sealing insert or the inlet hose
1)
must be installed at the inlet for safe operation. As delivered, the inlet port is sealed with a sealing in-
sert.
01.
To install the inlet hose
(5)
, unscrew the sealing inser
t
(2)
from the inlet por
t
(1)
(
A/F 10 mm).
02.
Screw the coupling
(3)
, part of the inlet hose
(5)
, into the inlet por
t
(1)
.
03.
Fasten the inlet hose
(5)
with coupling nut
(4)
to the c
oupling
(3)
.