19
Measurement
cal
CAL
INST. FACTOR
In narrow installation conditions, the conductivity
measurement is influenced by the sensor’s distance
to the wall and the wall material. This effect can be
compensated for by the installation factor. The meter
corrects the cell constant by multiplying it with the
installation factor. The value of the installation factor
depends on the diameter and the conductivity of the
pipe as well as on the sensor’s distance from the wall.
If the distance from the wall is sufficient (> 15 mm
(0.59"), DN 80 or larger), it is not necessary to consid-
er the installation factor (1.00). If the distance from
the wall is smaller, the installation factor increases
(> 1) when the pipe is electrically insulating and de-
creases (< 1) when the pipe is electrically conductive.
See the instructions in the sensor manufacturer’s
documentation.
cal
Value blinks
Use
to set the installation factor.
cal
Calibration is performed.
Display: CELL CONSTANT, INST. FACTOR.
The meter then automatically returns to measuring mode.
INST. FACTOR Calibration
(For inductive conductivity measurement only or with Memosens
4-electrode sensor with specification of installation factor)
Selected in the configuration menu.
Calibration
Calibration
Note:
To abort calibration, you can press
meas
at any time.