WARNING! Impairment of explosion protection.
Never use digital Memosens sensors without Ex approval in a hazardous location. For
these applications, you must use Memosens sensors with Ex approval. These sensors,
as well as the hazardous-area cables, are marked by an orange-red ring.
14
Connecting a Sensor
The Portavo 904(X) COND provides several connections so that many types of sen-
sors can be used for measurement (see illustration below).
Note that only
one
sensor may be connected to the meter at a time.
The meter automatically recognizes a connected Memosens sensor and switches
accordingly. Memosens is signaled in the display.
Separate Temperature Probe
Note:
Temperature measurement using a separate temperature probe is only possi-
ble when no Memosens sensor is connected.
After power-on, a separate temperature probe is automatically recognized. If you
want to replace the temperature probe, you must switch off the meter and then
switch it on again.
a
b
c
d
e
Connections
a - Micro USB port
b - M8, 4 pins for Memosens sensors
c - Temperature probe GND
d - Temperature probe
e - DIN socket, 8 pins for analog sensors
Commissioning
Memosens sensors have a cable coupling, which allows convenient replacement of
sensors while the cable remains connected to the meter. The connecting cable is
connected to socket
b
(M8, 4 pins for Memosens sensors).