18
Calibration
0.1 / 0.01 MOL KCL Calibration
(Automatic calibration with KCl solution)
The calibration method is selected in the configuration menu.
Measurement
cal
CAL
0.1/0.01 MOL KCL
PRESS CAL
Dip sensor in KCl solution.
The meter automatically compensates for the
temperature deviation!
cal
Measured value
Temperature
Conductivity KCl
Hourglass blinks
Calibration is performed.
Display: CELL CONSTANT.
The meter then automatically returns to measuring mode.
Calibration
Important notes:
• Make sure that the values of the calibration solutions used correspond exactly
to those specified in this manual.
If not, the resulting cell constant will be incorrect.
• When calibrating in a liquid, make sure that the conductivity sensor, any sepa-
rate temperature probe, and the calibration solution have the same tempera-
ture. Only this ensures that the cell constant is determined correctly.
Note:
To abort calibration, you can press
meas
at any time.