19
Calibration
Measurement
cal
CAL
TEMP. OFFSET
You can specify an offset for the temperature
measured by the sensor.
After calibration has been activated, the follow-
ing values are listed in the display:
• temperature setpoint
• temperature measured by sensor
• offset (display in K)
cal
Temperature setpoint value
blinks.
Use
to set the temperature setpoint value.
cal
Calibration is performed, the offset value is indicated.
The meter then automatically returns to measuring mode.
TEMP. OFFSET OFFSET (option)
Temperature calibration (offset)
Selected in the configuration menu.
Calibration
Note: To abort calibration, you can press meas at any time.