64
Commissioning
Connecting the
High-Voltage Cable
1. Shorten the cable support sleeve
(G)
corresponding to the diameter of
the high-voltage cable (HV cable).
2. Push the cable support sleeve onto the HV cable.
3. Strip the insulation from the HV cable as specified by the manufacturer of
the cable lug.
4. Crimp the ring cable lugs onto the ends of the HV cables
(use a tool approved by the manufacturer of the cable lug).
5. Connect the HV cables to the HV+ terminal
(E)
and the HV- terminal using
a washer and M5 nut for each terminal (recommended torque: 2 Nm).
Cut-off segment on the
cable support sleeve with
imprinted diameter
E
F
+
−
G
P50000 – Connecting the Input
CAUTION: Ensure correct guidance of the high-voltage cable!
Route the high-voltage cable as shown in the fig. on page 64.
Specifications of the high-voltage cable:
• pure copper wire
• cross-section: 1.5 mm
2
to 16 mm
2
e.g.,
Huber & Suhner
RADOX® 9 GKW-AX 3600V M single core
Overview of High-Voltage Input Terminals
Signal
Connection
Wire cross-section
Conditions/
Notes
HV+
M5 ring cable lug
Max. 16 mm
2
+ input
HV-
M5 ring cable lug
Max. 16 mm
2
– Input
WWW.OCTSOURCES.COM