60
Commissioning
Configuration
The measuring range (input) and the output parameters can be adjusted using
two rotary switches.
Note:
Only multi-
range models have
rotary switches.
Input
Output
The functions of the two rotary switches:
• Rotary switch S1 (input)
– Selecting the input range
• Rotary switch S2 (output)
– Switching between current and voltage output
– Selecting the output range
– Selecting the cutoff frequency
Input Rotary Switch, Standard Models
S1 P51*00
K11-M1M/11
P51*00
K11-M2M/11
P51*00
K11-M3M/11
P51*00
K11-M4M/11
0
30 mV
120 mV
2 V
20 V
1
50 mV
150 mV
3 V
30 V
2
60 mV
180 mV
4 V
40 V
3
90 mV
250 mV
5 V
50 V
4
100 mV
300 mV
6 V
60 V
5
30 mV + shunt
monitoring
120 mV + shunt
monitoring
7 V
70 V
6
50 mV + shunt
monitoring
150 mV + shunt
monitoring
8 V
80 V
7
60 mV + shunt
monitoring
180 mV + shunt
monitoring
9 V
90 V
8
90 mV + shunt
9
100 mV + shunt
monitoring
300 mV + shunt
monitoring
20 V
120 V
WWW.OCTSOURCES.COM