Installation and initial start-up
Operating Instructions LIQUIPORT EN
28
KNF Flodos | BA_LIQUIPORT_EN_11_153820.docx | Translation of original operating instructions, English
6.2.2 Connect hydraulics
Only connect components to the pump that are designed to han-
dle the hydraulic data of the pump (see Chapter 3).
Only use hoses that are suitable for the maximum operating pres-
sure of the pump (see Section 3).
Only use hoses that are sufficiently chemically resistant to the liq-
uids being transferred.
The connections described below apply to standard products. Dif-
ferent connections may apply for customer-specific projects (PML
or PL).
NOTE
Arrows on the pump head indicate the flow direction.
Use of hose connectors
1
Hose
2
Connector
3
Hose clip
Fig. 12: Hose connector ID 8mm or ID 12mm
Remove protective caps from connections.
Suction and pressure lines (LIQUIPORT 100 and LIQUIPORT
1.100: hose ID 8mm; LIQUIPORT 300 and LIQUIPORT 1.300:
hose ID 12mm), cut off straight with a sharp knife.
Push the hoses onto the connectors as far as they will go.
For pressure applications with LIQUIPORT 1.100 or LIQUIPORT
1.300, secure hoses with suitable hose clips.
Check that the hoses and transition joints (hose connector/hose)
are fitted correctly and securely.
Check that the system is leak-tight.
Connected
components
Hoses
Customer-specific
pumps (PL, PML)