Vacuum Degasser Unit
15
POWER LED
INIT / READY LED
Figure 7
Front view of Smartline Manager with Degasser LEDs
INIT
(Orange)
Indicates when the vacuum level is outside of the acceptable operating
range. Normally it will come on at initial power-up and remain lit during
pump down. It will go off in a few seconds when the vacuum level
goesbelow 100 mm of Hg (133 mbar) absolute. If an error occurs, this
LED will flash in one of two modes:
Flashing on and off in even 1-second intervals: pump was not able to
reach vacuum set point, indicating a possible leak in the system.
Flashing on for 1 second and off for 2 seconds indicates a vacuum signal
error. Please contact your Service.
READY
(Green)
Indicates when vacuum level is within acceptable operating range.
Normally it will come on after the initial pump-down, and remain on as
long as the Vacuum Degasser is powered up and vacuum level is below
100 mm of Hg absolute.
Rear Panel Connections
Connection with
Smartline Pump 1000
Degasser Exhaust
Figure 8
Rear view of Smartline Manager with Vacuum Degasser Unit
Exhaust Port
The gas pumped out of the vacuum chamber leaves the unit through the
exhaust port (see Figure 8). You can connect a gas tubing to the exhaust
port to avoid contaminating the laboratory with solvent gas molecules.