AZURA® Valve Unifier VU 4.1 Instructions V6855
Unpacking and setup 16
6.5.8
APIPA
To assign an IP address via APIPA configuration, both PC and device LAN
control must be configured to DHCP mode. If no DHCP server or router
can be detected, the device switches into APIPA configuration, and obtain
an IP address automatically. This process may take several minutes.
6.6
Remote control
On the rear of the valve drive are sockets on a terminal strip. Signals can
be send and received by other devices via those sockets. The signals are
for example start signals of a pump or detector which are connected to
the START IN connector. All voltages between GROUND and the corre-
sponding input or output must be connected.
The device has been switched off.
The power plug has been pulled.
Operating tool
Electronic defect
Connecting cables to the multi-pin connector of a switched on device
cau ses a short circuit.
Î
Turn off the device before connecting cables.
Î
Pull the power plug.
Electronic defect
Electrostatic discharge can destroy the electronics.
Î
Wear a protective bracelet against electrostatic discharge and ground.
1.
Insert the operating tool
3
into an upper small opening on
the front of the terminal strip
1
.
2.
Lead the cable into the opening
2
below the inserted operating
tool.
3.
Remove the operating tool.
Check if the cables are firmly atta-
ched. Push the terminal strip onto
the multi-pin connector. Finish the
installation. Put the device into
operation.
Prerequisites
Tools
Process
3
1
2
Next steps