AZURA® Valve Unifier VU 4.1 Instructions V6855
Operation 26
Menu
Explanation
Figure
Maximum
Pressure
Information of the maximum
pressure of the mounted valves
PMax
200 bar
Part Number
Part number of the mounted
valve
Part# EV
X35XXXX
7.5
Setting the valve position
Depending of the setting in Confirmation Mode, the valve position can
be set immediately (OFF) or after confirmation (ON).
7.5.1
Setting the valve position immediately
Process
Figure
1.
In main display, set the position via the navi-
gation buttons.
2.
Release the navigation button. The position is
set immediately.
POS
1 / 4
W
7.5.2
Setting the valve position after confirmation
Process
Figure
1.
Change to Confirmation Mode: Main Display
> Drive Setup > Confirmation Mode
2.
Press selection button.
3.
Use the navigation buttons to set the setting
from OFF to ON.
4.
Finish the process by pressing the confirma-
tion but ton.
Confirm
mode of
f
5.
Change to main display.
6.
Use the navigation buttons to set a value for
the posi tion.
7.
Confirm the selection by pressing the confir-
mation button.
POS
1 / 4
W