63
Technical data
AZURA® Detector DAD 6.1L, DAD 2.1L, MWD 2.1L Instructions, V6700
Detection
Wavelength
verifica tion
Internal holmium filter and deuterium lines
Noise
± 5 µAU at 254 nm
Drift
400 µAU/h at 254 nm
Linearity
> 1.6 AU at 274 nm, typically 2.5 AU
Time constants
0.0 / 0.1 / 0.2 / 0.5 / 1.0 / 2.0 / 5.0 / 10.0 s
Integration time
Automatic (5–1000 ms)
Communication
Interfaces
LAN (RJ-45), RS-232 (SUB-D 9), multi-pin
connector, analog (RCA cinch connector)
Control
Front panel, Mobile Control, software,
event con trol, analog, terminal protocol
Inputs
Error (IN), Start (IN), Autozero, Event 1–2
Outputs
Error (OUT), +5 V, Valve +24 V, Valve (OUT),
Start (OUT)
Analog output
1 x 0–5 V scalable, 20 bit, offset adjustable
Technical Parameters
GLP
Detailed report including lamp recognition,
ope rating hours, lamp operating hours,
number of lamp ignitions
Display
Mobile Control (optional)
Ambient conditions
Temperature range 4–40 °C, 39.2–104 °F
Humidity: below 90 %, non-condensing
General
Power supply
100–240 V, 50–60 Hz, 65 W
Dimensions
361 x 158 x 523 mm (H x W x D)
Weight
12.2 kg
Leak sensor
yes